SKU : PU-2089
Quantity
Short Description
Jasco PU-2080 Plus Quaternary Gradient solvent delivery part for performing of Gradient HPLC System. High Precision flow rate pump
Specifications
Pumping Method: SSQD (Super-Spacer Quick-Displacement) reciprocating system
Flow Rate Range: 1 µL/min to 10.0 mL/min (0.001–10 mL/min)
Flow Rate Precision: < 0.1% RSD (measured between 0.2 and 5.0 mL/min)
Maximum Pressure: 50 MPa (for flow rates < 5 mL/min)
35 MPa (for flow rates > 5 mL/min)
Solvent Mixing: Low-pressure quaternary gradient mixing for
up to 4 solvents (composition precision ±0.2%)
Programming: Supports time-based programming of flow, composition,
pressure, and relay events (up to 10 files, 64 steps per file)
Dimensions & Weight: 150 (W) x 470 (D) x 225 (H) mm
Power Requirements: AC 100–240V ±10%, 50/60 Hz, 185 VA
Additional Features: Built-in 4-channel vacuum degasser, ceramic plungers, and automatic plunger cleaning mechanism.